Trimble 4000 (Active RCI) - 11

This driver will be discontinued in a future QINSy release. Please contact QPS support if you need to make use of this driver in the future.


Description

Driver for Trimble Remote Control Interface (RCI) protocol.
Driver to be used to obtain raw GPS pseudo-ranges and various GPS broadcast data from a Trimble 4000 series GPS receiver. 

Driver Information

Driver Trimble 4000 (Active RCI)  Interface Type Serial  Driver Class Type Interrogated 
UTC Driver (question)  No Input / Output Input (two-way)  Executable  DrvTrimbleRCI.exe
Related Systems
Related Pages  



System Configuration

In configuring the Remote Control Interface (RCI) of a Trimble 4000 receiver use the CONTROL function key on the receiver.

The following description defines only the essential settings; those not shown are not important to the operation of QINSy:

RTCM

OFF

NMEA Output

OFF

Baud Rate Format

Baudrate

57600 (or as high as possible)

Databits

8

Parity

N

Stopbits

1

Mask/Sync Time

Mask

0

PDOPmask

99.9

SV synctime

1.0

Positioning Mode

Lat/Long/Hght or Lat/Long

(NB: this is one setting)

Cycle Printouts

OFF

Remote Protocol

4000 A/S compatible

GENERAL PROCEDURE FOR SETTING UP AND TESTING TRIMBLE RECEIVERS:

  1. Cycle through all the control functions and set the controls to the values defined above.
  2. Start the HyperTerminal program and set the required COM port to the correct I/O parameters. At this juncture no Trimble data should be visible on the screen.
  3. Turn on one of the cycle printouts, defining the port to which the QINSy system is connected.

    The ASCII position printout is useful for this purpose. Using the HyperTerminal program, check that you are receiving the position printout correctly. If no data is being read, check the interface settings and all the electrical connections.
  4. Turn off cycle printout, and, using HyperTerminal, check that no data is being received.
  5. Make sure the interface settings in DB Setup are set correctly, and start the Online module.
  6. After a short interval of perhaps 10-20 seconds, raw ranges should be displayed on the observation page of the Online module, if the connection is working correctly.

Interfacing Notes

This directive is applicable when interfacing the Trimble 4000 directly to the QINSy system, in the case where a Trimble OSM interface box is not supplied.
If a longer cable is needed, just make a direct 2-2, 3-3 and 7-7 connection from the DB25 end to the multiple COM board (e.g. Digi 8e).